FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3933275 · Received July 14, 2014

Report

Report Number
3004209178-2014-87042
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 15, 2014
Report Date
June 15, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH CURRENTS IN SPECIFICATION. NO UNEXPECTED WEAK BATTERY. NO BUTTON ERROR ALARM. INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE TO KEYPAD TRACE. MINOR SCRATCHES ON LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM AND A WEAK BATTERY ALARM. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 146 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409763 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 42 YR