FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3933149 · Received July 14, 2014

Report

Report Number
3004209178-2014-86927
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 15, 2014
Report Date
June 15, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO RAMPING NUMBERS ANOMALY NOTED. THE INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND CRACKED RESERVOIR TUBE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ON THE INSULIN PUMP WERE UNRESPONSIVE AND THE NUMBERS WERE RAMPING ON THEIR OWN. THE CUSTOMER'S BLOOD GLUCOSE WAS 140 MG/DL. THE CUSTOMER WAS ADVISED TO DISCONTINUE THE USE OF THE DEVICE AND TO REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.