FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3933117 · Received July 14, 2014

Report

Report Number
3004209178-2014-86904
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 14, 2014
Report Date
June 14, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, SELF AND BASIC OCCLUSION TEST. NO MOTOR ERROR ALARM NOTED. UNABLE TO PRIME DURING PRIME/A33 TEST DUE TO FAULTY FORCE SENSOR RESISTOR. UNABLE TO PERFORM THE OCCLUSION TEST AND EXCESSIVE NO DELIVERY TEST DUE TO PRIME/FILL ANOMALY. MOTOR PASSED MOTOR TEST. CRACKED RESERVOIR TUBE LIP NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING A BASAL, AND THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS. THE BLOOD GLUCOSE READING WAS 344 MG/DL; THE CUSTOMER'S MOTHER STATED THAT HE TREATED WITH A MANUAL INJECTION. THERE WERE NO SIGNIFICANT EVENTS LEADING TO THE ALARM OBSERVED. NO SYMPTOMS OF HIGH BLOOD GLUCOSE WERE REPORTED. UPON TROUBLESHOOTING, IT WAS FOUND THAT THE INSULIN PUMP WAS ABLE TO ADMINISTER INSULIN DURING A MANUAL PRIME. NO LEAK WAS FOUND, AND SETTINGS WERE FOUND TO BE CORRECT. TROUBLESHOOTING COULD NOT BE COMPLETED DUE TO THE CUSTOMER NOT HAVING THE TUBING CLAMP. ADVISED DISCONTINUATION OF THE DEVICE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410490 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 14 YR