FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3933098 · Received July 14, 2014

Report

Report Number
3004209178-2014-86898
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 14, 2014
Report Date
June 14, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD NO BUTTON RESPONSE ON UP ARROW BUTTON DUE TO CORRODED KEYPAD TRACES. CRACKED RESERVOIR TUBE LIP AND A MISSING END CAP STICKER NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD A KEYPAD ANOMALY AND SOME KEYS WERE UNRESPONSIVE. THE BLOOD GLUCOSE READING WAS 160 MG/DL. THE CUSTOMER STATED THAT THERE IS SOME FUNCTIONALITY ON THE ESC AND EASY BOLUS BUTTONS BUT THE OTHERS ARE NOT RESPONSIVE. ADVISED DISCONTINUATION OF THE DEVICE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410930 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 35 YR