FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 3933092 · Received July 14, 2014

Report

Report Number
3004209178-2014-86943
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 12, 2014
Report Date
June 14, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD MISSING SEGMENTS DUE TO CRACKED ZEBRA CONNECTOR ON LCD BOARD. THE INSULIN PUMP HAD A SMALL CRACK ON THE RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER CALLED REPORTING HIS INSULIN PUMP HAS AN INTERMITTENT PARTIAL DISPLAY. CUSTOMER'S BLOOD GLUCOSE LEVEL IS 123 MG/DL. CUSTOMER IS NOW USING AN OLD INSULIN PUMP AS A BACK UP. CUSTOMER USES ENERGIZER BATTERIES. THE DEVICE WAS NOT BUMPED OR DROPPED BEFORE THE ALARM. CUSTOMER WAS ADVISED THAT THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE OF IT AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410928 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 58 YR