530G INSULIN PUMP
Report
- Report Number
- 3004209178-2014-86943
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- June 12, 2014
- Report Date
- June 14, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP HAD MISSING SEGMENTS DUE TO CRACKED ZEBRA CONNECTOR ON LCD BOARD. THE INSULIN PUMP HAD A SMALL CRACK ON THE RESERVOIR TUBE LIP.
CUSTOMER CALLED REPORTING HIS INSULIN PUMP HAS AN INTERMITTENT PARTIAL DISPLAY. CUSTOMER'S BLOOD GLUCOSE LEVEL IS 123 MG/DL. CUSTOMER IS NOW USING AN OLD INSULIN PUMP AS A BACK UP. CUSTOMER USES ENERGIZER BATTERIES. THE DEVICE WAS NOT BUMPED OR DROPPED BEFORE THE ALARM. CUSTOMER WAS ADVISED THAT THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE OF IT AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 410928 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |