FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3932913 · Received July 14, 2014

Report

Report Number
1416980-2014-22570
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 19, 2014
Report Date
June 19, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS REPORTED TO BE AVAILABLE BUT HAS NOT YET BEEN RECEIVED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE REPORTED ISSUE. AN INTERNAL/EXTERNAL INSPECTION WAS PERFORMED AND THE DEVICE PASSED. FULL FUNCTIONAL TESTING, ELECTRICAL SAFETY TESTING, CALIBRATION TESTING, AND SIMULATED THERAPY WERE PERFORMED WITH NO ISSUES FOUND RELATED TO THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS THAT WERE RELATED TO THE REPORTED CONDITION. THE REPORTED CONDITION WAS NOT VERIFIED DURING EVALUATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE ISSUES WITH DRAINING ON THE HOME CHOICE (HC). THE CALLER STATED THE FIGURES DID NOT ADD UP TO THE ACTUAL FILL AMOUNT. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ARRANGED THE RETURN OF THE DEVICE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411242 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1