FDA Adverse Event Malfunction Summary report: N

SITTER II

MDR report key: 3932858 · Received March 11, 2014

Report

Report Number
2020362-2014-00074
Event Type
Malfunction
Date Received
March 11, 2014
Report Date
February 14, 2014
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION CODES: RESULTS: (OTHER) - EVALUATION OF THE RETURNED PRODUCT CONFIRMED THE REPORTED ISSUE. THE UNIT POWERS UP, BUT THE ALARM DOES NOT SOUND WHEN WEIGHT IS REMOVED FROM THE SENSOR PAD. THE RED LED LOW BATTERY INDICATOR WORKS AS IT SHOULD, BUT , THE FAIL SAFE DOES NOT WORK. THE NURSE CALL LIGHT TURNS OFF INTERMITTENTLY AT THE NURSE CALL TEST FIXTURE WHEN THE NURSE CALL TEST FIXTURE WHEN THE NURSE CALL CABLE IS WIGGLED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE ALARM HAS POWER, BUT DOES NOT SOUND WHEN WEIGHT IS REMOVED FROM THE SENSOR. ALSO, THE LED LIGHT DOES NOT LIGHT UP. THE DATE WHEN THE ISSUE WAS DISCOVERED IS UNK. NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
146337 SITTER II KMI J. T. POSEY CO. 8281 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK BED SENSOR PAD: MODEL 8307, LOT # UNK