FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 3932827 · Received July 14, 2014

Report

Report Number
2024168-2014-04505
Event Type
Injury
Date Received
July 14, 2014
Date of Event
June 17, 2014
Report Date
June 19, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED DEVICE OPERATES DIFFERENTLY THAN EXPECTED (NEEDLE TO CUFF MISS) WAS CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 6F SHEATH AFTER A CORONARY INTERVENTIONAL PROCEDURE. REPORTEDLY, A POSTERIOR CUFF MISS OCCURRED. HEMOSTASIS WAS ACHIEVED USING A SECOND PROGLIDE DEVICE. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410873 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 40421K1

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention SHEATH: 6F, BIVALIRUDIN