FDA Adverse Event Malfunction Summary report: N

2.4MM TI VA-LCP 2-COL DSTL RAD PL NRW 6H HD/3H SHAFT/LT-STER

MDR report key: 3932776 · Received July 14, 2014

Report

Report Number
1000562954-2014-10133
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
SYNTHES MEZZOVICO
Product Code
HRS
PMA / PMN Number
PK092556
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL COMMON DEVICE NAME HWC. NONCONFORMANCE REPORTS (NCR (B)(4)) IS MENTIONED IN THE MANUFACTURING DOCUMENT, INDICATING THAT PART 04.111.531S / LOT 7960446 DROPPED DURING TRANSPORT. A CAUSAL RELATIONSHIP BETWEEN NCR (B)(4) AND THE CONDITION DESCRIBED IN (B)(4) IS ASSESSED AS VERY UNLIKELY. IT CAN BE CONCLUDED THAT THE REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED DURING A SMITH FRACTURE CASE (LEFT DISTAL RADIUS PART) THE SURGEON MOUNTED A GUIDING BLOCK TO THE PLATE AND ATTACHED THE DRILL SLEEVE, BUT IT DID NOT FIT IN PROPERLY. THEREFORE THE SURGEON INSERTED THE LOCKING SCREW BY HAND, THE SCREW DID NOT LOCK AND WENT THROUGH THE GUIDING BLOCK. SINCE OTHER SCREWS WERE INSERTED PROPERLY, THE SURGEON DID NOT REMOVE THE SCREW IN QUESTION. AFTER CHECKING THAT THE SCREW HAD NOT GONE THROUGH THE BACK SIDE, THE SURGEON COMPLETED THE OPERATION. THIS RESULTED IN A 10 MINUTE DELAY AND RESIDUE IN THE PATIENT. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409152 2.4MM TI VA-LCP 2-COL DSTL RAD PL NRW 6H HD/3H SHAFT/LT-STER PLATE,FIXATION,BONE HRS SYNTHES MEZZOVICO 7960446

Patients

Seq Age Sex Outcome Treatment
1 85 YR