FDA Adverse Event Malfunction Summary report: N

OPTICROSS?

MDR report key: 3932766 · Received July 14, 2014

Report

Report Number
2134265-2014-04011
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
April 10, 2014
Report Date
June 16, 2014
Manufacturer
BOSTON SCIENTIFIC - FREMONT (SUD)
Product Code
OBJ
PMA / PMN Number
K123621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS RECEIVED FOR EVALUATION. EVALUATION OF THE RETURNED DEVICE REVEALED THAT THERE WAS AN OPEN HOLE WAS OBSERVED AT THE SHEATH LAP JOINT SECTION OF THE DEVICE. FLUID WAS LEAKING FROM THE OPEN HOLE AT THE SHEATH LAP JOINT ASSEMBLY WHEN THE CATHETER WAS FLUSHED. THE TELESCOPE ASSEMBLY WAS NOT ABLE TO PROPERLY PULL BACK, ADVANCE, OR RETRACT. SINCE THE TELESCOPE CANNOT ADVANCE THE TRANSDUCER DISTAL HOUSING (TDH) TO THE MOST DISTAL POSITION, THE DISTANCE FROM THE DISTAL END OF THE TRANSDUCER HOUSING TO THE TIP OF THE CATHETER WAS NOT MEASURED. DURING IMAGE CHARACTERIZATION TESTING, NO IMAGE APPEARED IN THE SYSTEM DUE TO ELECTRICAL OPEN AT PROXIMAL. THE IMAGING CORE WAS FOUND BROKEN OFF FROM THE ROTATORY ASSEMBLY¿S CONNECTOR SHAFT. THE IMAGING CORE WAS STILL WITHIN THE TELESCOPE SECTION OF THE DEVICE. IMAGING CORE WINDUP WAS FOUND WITHIN THE TELESCOPE SECTION OF THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(4) 2014. IT WAS REPORTED THAT LOST IMAGE OCCURRED. THE 99% STENOSED LESION WAS SEVERELY TORTUOUS AND SEVERELY CALCIFIED. DURING THE PROCEDURE, AN OPTICROSS IMAGING CATHETER WAS USED DURING INTRAVASCULAR ULTRASOUND (IVUS); HOWEVER, LOST IMAGE OCCURRED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CONDITION IS GOOD. HOWEVER, DEVICE ANALYSIS REVEALED AN OPEN HOLE AT THE SHEATH LAP JOINT SECTION OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409863 OPTICROSS? CATHETER, ULTRASOUND, INTRAVASCULAR OBJ BOSTON SCIENTIFIC - FREMONT (SUD) H749518080 16766612

Patients

Seq Age Sex Outcome Treatment
1