PROMUS ELEMENT ?
Report
- Report Number
- 2134265-2014-04037
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- January 15, 2014
- Report Date
- June 17, 2014
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR.: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE FOUND THE DEVICE WAS RETURNED TO THE COMPLAINT INVESTIGATION SITE WITH THE STENT DETACHED FROM THE BALLOON. THE BALLOON WAS FOUND TO HAVE THE STENT IMPRESSIONS ON IT AND PILLOWING OF THE BALLOON INDICATING THAT THE STENT WAS CRIMPED PER PROCESS IN THE CORRECT LOCATION DURING MANUFACTURING. AN EXAMINATION FOUND NO DEFECTS ON THE BALLOON AND TIP SECTIONS OF THE DEVICE. THERE WERE VARIOUS HYPOTUBE KINKS ALONG THE LENGTH OF THE UNIT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE WAS UNABLE TO BE DETERMINED. (B)(4).
REPORTABLE BASED ON DEVICE ANALYSIS APPROVED ON (B)(4) 2014. IT WAS REPORTED THAT CROSSING DIFFICULTIES WAS ENCOUNTERED. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE 100% STENOSED, 20X2.5MM TARGET LESION CONTAINED A >=90 DEGREE BEND AND WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED RIGHT CORONARY ARTERY (RCA). A 2.50X20MM PROMUS ELEMENT DRUG-ELUTING STENT WAS SELECTED FOR USE AND ADVANCED BUT FAILED TO CROSS THE LESION. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED STENT DETACHMENT/SEPARATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 409859 | PROMUS ELEMENT ? | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911320250 | 15883988 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |