FDA Adverse Event Malfunction Summary report: N

NC TREK CORONARY DILATATION CATHETER

MDR report key: 3932638 · Received July 14, 2014

Report

Report Number
2024168-2014-04499
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 17, 2014
Report Date
June 19, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K110134
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED BALLOON RUPTURE WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE COMPLAINT HANDLING DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT USING A RADIAL ARTERY ACCESS APPROACH DURING A PROCEDURE OF THE NON-CALCIFIED, NON-TORTUOUS, DE NOVO, MID CIRCUMFLEX ARTERY DURING THE FIRST INFLATION OF THE 2.25 X 15 MM NC TREK BALLOON DILATATION CATHETER (BDC) AT 8 ATMOSPHERE (ATM) THE BALLOON FAILED TO HOLD PRESSURE AND A LEAK AT THE PROXIMAL PORTION OF THE BALLOON WAS NOTED. THE BDC WAS WITHDRAWN FROM THE PATIENT ANATOMY WITHOUT ANY REPORTED ISSUES. A SIMILAR SIZED BDC WAS USED TO SUCCESSFULLY COMPLETE THE CASE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409569 NC TREK CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 40205G1

Patients

Seq Age Sex Outcome Treatment
1 INFLATION: ABBOTT INDEFLATOR