FDA Adverse Event Injury Summary report: N

ASCENDRA 3 INTRODUCER SHEATH SET

MDR report key: 3932607 · Received July 14, 2014

Report

Report Number
2015691-2014-01582
Event Type
Injury
Date Received
July 14, 2014
Date of Event
July 9, 2014
Report Date
July 9, 2014
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPT
PMA / PMN Number
P110021
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PER THE INSTRUCTIONS FOR USE (IFU), ACCESS SITE COMPLICATIONS/CARDIOVASCULAR INJURY, INCLUDING PERFORATION OF THE VENTRICLE OR MYOCARDIUM, BLEEDING, AND INJURY AT THE SITE OF VENTRICULAR ACCESS THAT MAY REQUIRE REPAIR ARE POTENTIAL COMPLICATIONS ASSOCIATED WITH THE TRANSAPICAL TAVR PROCEDURE. PER LITERATURE REVIEW, RISK FACTORS FOR APICAL LACERATION AND BLEEDING INCLUDE FRIABLE TISSUE, FATTY APEX, CHRONIC STEROID USE, DILATED LV WITH THINNED WALLS, AND HYPERTENSION DURING REMOVAL OF THE SHEATH. WHILE PATIENT CHARACTERISTICS ARE IMPORTANT, ACHIEVING GOOD HEMOSTATIC CONTROL OF THE LV APEX IS ONE OF THE MOST CRITICAL STEPS IN ENSURING THE SUCCESS OF THE TRANSAPICAL PROCEDURE, PARTICULARLY IN THE ELDERLY WITH FRIABLE TISSUE. ADDITIONAL LITERATURE REVIEW CONFIRMS THERE IS A HIGHER INCIDENCE OF APICAL BLEEDING IN FEMALE PATIENTS AND PATIENTS OVER (B)(6). THIS INFORMATION CORRELATES WITH REVIEW OF COMPLAINT HISTORY, REVEALING THAT THE MAJORITY OF APICAL BLEEDING COMPLICATIONS APPEAR TO BE RELATED TO SURGICAL TECHNIQUE AND/OR DISEASED VENTRICLES DURING THE INSERTION OR REMOVAL OF THE SHEATH. THE THV TRAINING MANUALS NOTE THAT REMOVAL OF THE SHEATH AND ATTEMPTED CLOSURE OF THE APICAL INCISION MAY BE ASSOCIATED WITH BLOOD LOSS. RAPID BURST PACING CAN BE USED TO LOWER THE SYSTEMIC BLOOD PRESSURE DURING SHEATH REMOVAL AND APICAL CLOSURE. THE THV TRAINING MANUALS ALSO RECOMMEND THE PHYSICIAN CONSIDER PERFORMING THE PROCEDURE UNDER FULL CPB SUPPORT FOR CERTAIN PATIENTS, INCLUDING THOSE WITH CARDIOGENIC SHOCK (CARDIAC INDEX < 2 L/MIN PER SQUARE METER) DESPITE VOLUME CHALLENGE AND INOTROPIC SUPPORT, PROFOUND LV, RV, OR BIVENTRICULAR DYSFUNCTION, AND NOTABLY FRIABLE APEX. IN THIS CASE, IT APPEARS THAT PATIENT FACTORS (PATIENT¿S AGE, GENDER, AND FRIABLE APEX TISSUE) CAUSED OR CONTRIBUTED TO THIS EVENT. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING A TRANSAPICAL TAVR PROCEDURE, THE SHEATH WAS INSERTED INTO THE APEX AND WHEN THE PHYSICIAN WENT TO TIGHTEN THE PURSE STRINGS, THE PATIENT'S TISSUE TORE AND WOULD NOT HOLD THE SUTURES. THE PATIENT WAS PLACED ON PUMP AND THE PATIENT'S APEX WAS REPAIRED WITH SUTURES. THE PROCEDURE WAS ABORTED, THE PATIENT WAS TAKEN OFF PUMP AND TRANSFERRED TO THE UNIT IN STABLE CONDITION. THERE WAS NOTHING ABNORMAL NOTICED WITH THE SHEATH PRIOR TO USE. THE APEX TEAR WAS ATTRIBUTED TO THE PATIENT¿S FRIABLE TISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409284 ASCENDRA 3 INTRODUCER SHEATH SET AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES 9120IS 59627490

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention