FDA Adverse Event Malfunction Summary report: N

5MM MONOPOLAR HANDLE 33CM

MDR report key: 3932593 · Received July 14, 2014

Report

Report Number
0002936485-2014-00495
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DURING THE INSPECTION OF THE 5MM MONOPOLAR HANDLE 33CM IT WAS CONFIRMED, THE INSULATION IS COMPROMISED. VISIBLE DINGS AND SCRATCHES WERE SEEN THROUGHOUT THE SHAFT. THE 5MM MONOPOLAR HANDLE 33CM PASSED THE INSULSCAN TEST. THE PROBABLE ROOT CAUSE COULD BE USER MISHANDLING AND OR WEAR AND TEAR. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409242 5MM MONOPOLAR HANDLE 33CM ENDOSCOPE AND/OR ACCESSORIES KOG STRYKER ENDOSCOPY-SAN JOSE 06202014

Patients

Seq Age Sex Outcome Treatment
1