COYOTE? ES
Report
- Report Number
- 2134265-2014-03983
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- June 13, 2014
- Report Date
- June 16, 2014
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- LIT
- PMA / PMN Number
- K080982
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MFR. THE CATHETER WAS TACTILELY AND VISUALLY INSPECTED; THE SHAFT WAS KINKED SLIGHTLY IN TWO PLACES 78CM FROM THE DISTAL TIP. THE BALLOON AND SHAFT WERE INSPECTED UNDER MAGNIFICATION. THERE WAS A BALLOON PINHOLE OVER THE MARKERBAND; THE INSPECTION PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. INSPECTION OF THE REMAINDER OF THE DEVICE PRESENTED NO DAMAGE OR IRREGULARITIES. THERE WAS NO EVIDENCE OF ANY PRODUCT QUALITY DEFICIENCIES CONTRIBUTING TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS LEFT ANTERIOR TIBIAL ARTERY. A 1.5MM X 20MM X 142CM COYOTE ES BALLOON CATHETER WAS ADVANCED TO DILATE THE LESION. DURING PROCEDURE THE BALLOON WAS INFLATED ONCE AT 6 ATMOSPHERES AND CONTRAST DYE LEAKED WAS FOUND FROM THE COMPLAINT DEVICE. THE DEVICE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 409485 | COYOTE? ES | CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL | LIT | BOSTON SCIENTIFIC - MAPLE GROVE | H74939134152010 | 16851408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INTRODUCER SHEATH: 6FR SHEATH (TERUMO)| GUIDE WIRE: AGOSAL 014 (ST.JUDE MEICAL) |