FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 3932302 · Received April 4, 2014

Report

Report Number
3006697241-2014-00309
Event Type
Malfunction
Date Received
April 4, 2014
Date of Event
March 6, 2014
Report Date
March 6, 2014
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CASTERS ARE WORN, MOST LIKELY DUE TO NORMAL WEAR AND TEAR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED ALL BRAKE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BRAKES WILL SET BUT DO NOT HOLD. THE STRETCHER WAS LOCATED IN THE BED SHOP AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
205632 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILLROM DE MEXICO S DE RL DE CV 8000

Patients

Seq Age Sex Outcome Treatment
1