FDA Adverse Event Malfunction Summary report: N

TI CLICK'X® LOCKING CAP FOR TI 3-D HEAD

MDR report key: 3932267 · Received July 14, 2014

Report

Report Number
1000562954-2014-10120
Event Type
Malfunction
Date Received
July 14, 2014
Report Date
June 16, 2014
Manufacturer
SYNTHES MEZZOVICO
Product Code
NKB
PMA / PMN Number
PK082572
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL COMMON DEVICE NAMES: MNI, MNH, KWP, KWQ. NO NONCONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: A PD EVENT EVALUATION WAS CONDUCTED. THE REPORT INDICATES THAT THE SITUATION AS DESCRIBED IN THE COMPLAINT IN (B)(6) COULD NOT BE REPLICATED. THE DAMAGE OF THE HEX DRIVE OF THE LOCKING CAPS, WHICH INDICATES CLOCK WISE AND COUNTER CLOCK WISE LOADING OF THE SET SCREWS, MIGHT BE A SIGN THAT WRONG HANDLING HAS BEEN THE CAUSE OF THE CLOGGING OF THE SET SCREWS. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED DURING PREPARATION, PRIOR TO SURGERY, THE NURSE NOTICED THAT THE CLICK' X CAPS WERE BLOCKED. THE DEVICES WERE NOT USED FOR THE SURGERY. THE EVENT DID NOT PROLONG SURGERY. THIS IS REPORT 2 OF 10 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410432 TI CLICK'X® LOCKING CAP FOR TI 3-D HEAD ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE NKB SYNTHES MEZZOVICO 8904771

Patients

Seq Age Sex Outcome Treatment
1