FDA Adverse Event Malfunction Summary report: N

TOTAL CARE BARIATRIC BED

MDR report key: 3932260 · Received April 4, 2014

Report

Report Number
1824206-2014-01096
Event Type
Malfunction
Date Received
April 4, 2014
Date of Event
March 8, 2014
Report Date
March 8, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE NURSE WAS NOT PUSHING THE SIDERAIL UP HARD ENOUGH TO ALLOW THE LATCH TO GRAB ONTO THE BED AND LATCH COMPLETELY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN INSERVICED THE NURSE HOW TO LATCH THE SIDERAIL TO RESOLVE THE ISSUE. THE BED SIDERAILS FUNCTIONED AS DESIGNED. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT THE RIGHT SIDERAIL WOULD NOT LATCH IN THE UP POSITION. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
205751 TOTAL CARE BARIATRIC BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1