ULTRASITE VALVE NEEDLELESS SYSTEM
Report
- Report Number
- 2523676-2014-00088
- Event Type
- Malfunction
- Date Received
- April 4, 2014
- Date of Event
- February 27, 2014
- Report Date
- March 11, 2014
- Manufacturer
- B. BRAUN MEDICAL, INC.
- Product Code
- FPA
- PMA / PMN Number
- K031923
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B BRAUN MEDICAL INC. (B)(4). ONE USED ULTRASITE VALVE, WITHOUT PACKAGING, WAS RECEIVED FOR EVAL. THE HUBER NEEDLE TUBING, THAT WAS REPORTED TO BE CONNECTED TO THE VALVE, WAS NOT RETURNED. NO ABNORMALITIES OR ANOMALIES WERE VISUALLY OBSERVED ON THE ULTRASITE VALVE. THE VALVE WAS THEN SUBJECTED TO FLOW AND WATER PRESSURE (LEAKAGE) TESTING ACCORDING TO SPECIFICATION WITH ACCEPTABLE RESULTS. THERE WAS NO LEAKAGES OBSERVED WITHIN THE VALVE OR AT THE LUER CONNECTORS DURING THE TESTING TIME FRAME. THE RETURNED SAMPLE MET REQUIREMENTS ACCORDING TO SPECIFICATION, AND THE REPORTED FAILURE COULD NOT BE REPRODUCED. THE REPORTING FACILITY DID INDICATE THAT THE NURSE MAY NOT HAVE TIGHTENED THE ULTRASITE. IT IS INDICATED ON THE INSTRUCTIONS FOR USE (IFU) FOR THE REPORTED PRODUCT CATALOG NUMBER TO "FIRMLY ATTACH INJECTION SITE TO DESIRED CONNECTION." NO ADVERSE QUALITY TRENDS OF THIS NATURE WERE IDENTIFIED DURING THE COMPLAINT REVIEW PROCESS FOR THE REPORTED CATALOG NUMBER. IF ADD'L PERTINENT INFO BECOMES AVAILABLE, A F/U REPORT WILL BE FILED.
AS REPORTED BY THE USER FACILITY: REPORTS DURING INFUSION OF BICARB AND CHEMO, LEAKAGE OCCURRED AT THE CONNECTION BETWEEN THE HUBER NEEDLE TUBING AND ULTRASITE VALVE CAUSING A CHEMO SPILL. THE RN MANAGER STATED THE NURSE MAY NOT HAVE TIGHTENED THE ULTRASITE AND IT BEGAN LEAKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 205498 | ULTRASITE VALVE NEEDLELESS SYSTEM | ULTRASITE VALVE | FPA | B. BRAUN MEDICAL, INC. | CSU100 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | HUBER NEEDLE TUBING |