FDA Adverse Event Injury Summary report: N

EVIA DR-T

MDR report key: 3932233 · Received July 14, 2014

Report

Report Number
1028232-2014-002397
Event Type
Injury
Date Received
July 14, 2014
Date of Event
May 23, 2014
Report Date
July 1, 2014
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
DXY
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THEREFORE, THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE WERE RE-INVESTIGATED. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING PRODUCTION AND FINAL ACCEPTANCE TEST. THE BIOTRONIK STERILIZATION PROTOCOL CONFIRMED THAT ALL STERILIZATION PARAMETERS, SUCH AS GAS CONCENTRATION, TEMPERATURE, HUMIDITY, ETC., WERE IN THEIR SPECIFIED RANGES. ALSO, THE INVESTIGATION OF THE MICROBIOLOGICAL INDICATORS SHOWED THE SUCCESSFUL COMPLETION OF THE STERILIZATION PROCESS. IN SUMMARY, THE INFECTION WAS NOT DEVICE RELATED.

Description of Event or Problem · 1

THIS SYSTEM WAS REMOVED DUE TO SYSTEMIC INFECTION. THERE IS NO INDICATION THAT THE SYSTEM WAS REPLACED. THE HOSPITAL RETAINED THE DEVICES. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411067 EVIA DR-T PACEMAKER DXY BIOTRONIK SE & CO. KG 359529

Patients

Seq Age Sex Outcome Treatment
1 94 YR Hospitalization