LUMAX 540 DR-T
Report
- Report Number
- 1028232-2014-002389
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- June 4, 2014
- Report Date
- July 2, 2014
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- MRM
- PMA / PMN Number
- P000009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT THE ICD WAS SUBJECTED TO AN ELECTRICAL ANALYSIS. THEREBY THE CLINICAL OBSERVATION WAS CONFIRMED, THE DEVICE WAS NOT INTERROGATABLE. THE DEVICE HISTORY RECORD WAS INSPECTED, DOCUMENTING THAT THE DEVICE PERFORMED AS EXPECTED DURING THE DEVICE MANUFACTURING. THE ICD WAS OPENED. THE VISUAL INSPECTION OF THE INNER ASSEMBLY SHOWED NO ANOMALIES. THE BATTERY WAS FOUND TO BE DEPLETED. THE ELECTRONIC MODULE WAS ATTACHED TO AN EXTERNAL POWER SUPPLY. THE ELECTRICAL PARAMETERS, PARTICULARLY THE CURRENT CONSUMPTION OF THE ELECTRONIC MODULE WERE FOUND TO BE WITHIN SPECIFICATION WITH AN EXTERNAL POWER SUPPLY ATTACHED. THE ABILITY OF THE ELECTRONIC MODULE TO DELIVER THERAPIES WAS VERIFIED. THE ANTI-BRADYCARDIA PACING PULSES PROVED TO BE FLAWLESS IN AMPLITUDE AND FREQUENCY AS PROGRAMMED. A¿FIBRILLATION SIGNAL WAS APPLIED AND THE DEVICE DELIVERED A DEFIBRILLATION SHOCK AS SPECIFIED, DOCUMENTING A CORRECT SENSING AND SHOCK DELIVERY. IN PARTICULAR, THE SPECIFIED ENERGY LEVEL WAS REACHED. THE BATTERY WAS SENT TO THE MANUFACTURER FOR FURTHER ANALYSIS. ANALYSIS OF THE BATTERY THE MANUFACTURING RECORDS WERE INSPECTED, DOCUMENTING THAT THE BATTERY PARAMETERS WERE WITHIN SPECIFICATION DURING THE BATTERY MANUFACTURING. NO ANOMALIES WERE DOCUMENTED DURING THE PRODUCTION PROCESS, ASSOCIATED TO THIS BATTERY. THE VISUAL INSPECTION OF THE BATTERY DID NOT REVEAL ANY EXTERNAL SIGNS OF DAMAGE. THE VOLTAGE MEASUREMENT CONFIRMED THE BATTERY DEPLETION. NEXT, THE BATTERY WAS OPENED FOR DESTRUCTIVE ANALYSIS. DURING THE INSPECTION OF THE INNER ASSEMBLY AN INSULATION DAMAGE WAS IDENTIFIED, WHICH LED TO AN INCREASED INTERNAL SELF-DEPLETION WITHIN THE BATTERY AND THEREFORE TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS IMPLANTED FOR 55 MONTHS. THE CLINICAL OBSERVATION RESULTED FROM AN INTERNAL SELF-DEPLETION WITHIN THE BATTERY.
OUS MDR - AFTER AN IMPLANTATION PERIOD OF ABOUT 56 MONTHS, IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED DURING A ROUTINE FOLLOW-UP. THE PATIENT WAS ADMITTED TO HOSPITAL AND THE DEVICE WAS REPLACED. NO ADVERSE PATIENT SIDE EFFECTS HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 410963 | LUMAX 540 DR-T | ICD | MRM | BIOTRONIK SE & CO. KG | 360346 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |