FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 3932210 · Received July 14, 2014

Report

Report Number
1045834-2014-12422
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 26, 2014
Report Date
June 27, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE MET ALL MANUFACTURER'S SPECIFICATIONS. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED. AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNSPECIFIED SPINE SURGERY, IT WAS OBSERVED THAT THE MOTOR DEVICE WAS WOBBLING WHILE IN USE. IT WAS FURTHER REPORTED THAT THE DEVICE MADE THE BURR DEVICE WOBBLE DURING SURGERY. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH AN IDENTICAL SPARE DEVICE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE REPORTER INDICATED THAT THE PATIENT WAS ¿FINE¿. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409625 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 BURR DEVICE