ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2014-19857
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Report Date
- June 28, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PATIENT
Narratives
THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
FOLLOW-UP # 1 DATE OF SUBMISSION 07/31/2014-DEVICE EVALUATION: A RETAINED SAMPLE OF THE DEVICE HAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A RETAIN SAMPLE FROM THE SAME LOT NUMBER WAS TESTED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FORCE TEST, FILL TEST WAS PERFORMED WITH NO FAILURES OBSERVED OR FLUID OBSERVED LEAKING OUT AT THE PLUNGER END OF THE CARTRIDGE. THE CARTRIDGE FORCE READINGS WERE CONFIRMED TO BE WITHIN SPECIFICATIONS; NO FAILURES WERE NOTED RELATED TO THE LOSS OF PRIME COMPLAINT. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE WITH THE CARTRIDGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 409898 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | U200018 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |