FDA Adverse Event
Malfunction
Summary report: N
CLINITRON RITEHITE BED
MDR report key: 3932190
·
Received April 3, 2014
Report
- Report Number
- 1824206-2014-01089
- Event Type
- Malfunction
- Date Received
- April 3, 2014
- Date of Event
- March 7, 2014
- Report Date
- March 7, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- INX
- PMA / PMN Number
- K964223
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-R0M TECHNICIAN FOUND THE BRAKE CASTER NOT HOLDING DUE TO THE BRAKE PAD HAD WORN DOWN DUE TO NORMAL WEAR AND TEAR. THE TECHNICIAN REPLACED THE BRAKE CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED WAS MOVING WITH THE BRAKE ENGAGED. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200534 | CLINITRON RITEHITE BED | BED, AIR FLUIDIZED | INX | HILL-ROM, INC. | 0800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |