FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC BED

MDR report key: 3932172 · Received April 3, 2014

Report

Report Number
1824206-2014-01090
Event Type
Malfunction
Date Received
April 3, 2014
Date of Event
March 7, 2014
Report Date
March 7, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND DURING A BETWEEN PATIENT INSPECTION THE LEFT REAR, RIGHT FRONT, AND THE RIGHT REAR BRAKE CASTERS WOULD NOT HOLD AS THEY WERE OUT OF ADJUSTMENT FROM NORMAL WEAR AND TEAR. THE TECHNICIAN ADJUSTED THE REAR BRAKE CASTERS AND THE RIGHT FRONT BRAKE CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN REPORTED THE REAR BRAKE CASTERS AND THE RIGHT FRONT BRAKE CASTERS WOULD NOT HOLD. THE BED WAS LOCATED IN THE HILL-ROM SERVICE CENTER. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).