FDA Adverse Event Malfunction Summary report: N

PROCEDURAL STRETCHER

MDR report key: 3932166 · Received April 3, 2014

Report

Report Number
3006697241-2014-00303
Event Type
Malfunction
Date Received
April 3, 2014
Date of Event
March 4, 2014
Report Date
March 4, 2014
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE SIDE RAIL LATCH PLATE IS BENT. THE MOST LIKELY CAUSE OF A BENT LATCH PLATE IS FORCE SUCH AS THE SIDE RAIL WAS RAN INTO AN OBJECT DURING TRANSPORT. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2013 AND 2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE SIDE RAIL LATCH TO RESOLVE THE ISSUE. BASED ON TIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE SIDE RAIL WILL NOT LATCH. THE STRETCHER IS LOCATED IN THE HALL AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200647 PROCEDURAL STRETCHER WHEELED STRETCHER FPO HILLROM DE MEXICO S DE RL DE CV 8005

Patients

Seq Age Sex Outcome Treatment
1