5MM, 33CM PEEK MONOPOLAR HANDLE 33CM
Report
- Report Number
- 0002936485-2014-00486
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE ORACLE SERVICE RECORD CONFIRMS THE DEVICE HAS BEEN REPAIRED BEFORE AN INVESTIGATOR WAS ABLE TO PERFORM A FULL EVALUATION; THEREFORE THIS COMPLAINT INVESTIGATION WILL NOW BE CLOSED BASED ON THE REPAIR FINDINGS, AND PROBABLE ROOT CAUSE(S). THE PROBABLE ROOT CAUSE FOR THE REPORTED FAILURE INVOLVING THIS DEVICE COULD HAVE BEEN CAUSED BY: USER MISUSE, IMPROPER STERILIZATION METHODS, OR NORMAL WEAR THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED THROUGH THE SERVICE REPAIR. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 409819 | 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 06232014 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |