FDA Adverse Event Malfunction Summary report: N

SPECTURM INFUSION PUMP

MDR report key: 3932029 · Received February 1, 2014

Report

Report Number
1314492-2014-07259
Event Type
Malfunction
Date Received
February 1, 2014
Date of Event
January 1, 2014
Report Date
January 14, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE REPORTED SYMPTOM SYSTEM ERROR 322 CAUSED BY LOOSE SCREWS. THE UPPER AUXILIARY ASSEMBLY WAS REPLACED SINCE IT CONTAINS THE SCREWS. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FORM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69077 SPECTURM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1