FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3931962 · Received February 1, 2014

Report

Report Number
1314492-2014-07231
Event Type
Malfunction
Date Received
February 1, 2014
Date of Event
January 1, 2014
Report Date
January 14, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE REPORTED "NOT SEEING CLOSED DOOR" CAUSED BY A FAILED UPPER HOOK SWITCH. THE UPPER AUXILIARY ASSEMBLY WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A DEVICE WAS NOT RECOGNIZING A CLOSED DOOR. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.