FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC PLUS BED

MDR report key: 3931759 · Received April 15, 2014

Report

Report Number
1824206-2014-01198
Event Type
Malfunction
Date Received
April 15, 2014
Date of Event
March 18, 2014
Report Date
March 18, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND DURING A BETWEEN PATIENT INSPECTION THAT THE BED WAS GOING DOWN ON ITS OWN DUE TO THE LEFT SIDE RAIL COMMUNICATION CABLE MALFUNCTIONING. THIS WAS MOST LIKELY DUE TO NORMAL WEAR AND TEAR. THE LEFT SIDE RAIL COMMUNICATION CABLE WAS REPLACED TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN REPORTED THAT THE BED WAS MOVING DOWN ON ITS OWN. THE BED WAS LOCATED IN THE HILL-ROM SERVICE CENTER. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT NUMBER (B)(4).