ADVANTA BED
Report
- Report Number
- 1824206-2014-01193
- Event Type
- Malfunction
- Date Received
- April 15, 2014
- Date of Event
- March 18, 2014
- Report Date
- March 18, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
THE HILL-ROM TECHNICAL SUPPORT FOUND THE NURSE CALL WAS MOST LIKELY NOT WORKING DUE TO THE UNIVERSAL TELEVISION POWER CONTROL BOARD HAVING A RELAY CLICK. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE HILL-ROM TECHNICAL SUPPORT ADVISED THE CUSTOMER TO SWITCH UNIVERSAL TELEVISION POWER CONTROL BOARD TO RESOLVE THE ISSUE. THREE ATTEMPTS HAVE BEEN MADE REGARDING A RESOLUTION TO THIS CONTACT LINE, WITH NO RESPONSE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM TECHNICAL SUPPORT RECEIVED A REPORT FROM THE ACCOUNT STATING THE NURSE CALL WAS NOT WORKING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230998 | ADVANTA BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |