FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 3931737 · Received April 15, 2014

Report

Report Number
1824206-2014-01193
Event Type
Malfunction
Date Received
April 15, 2014
Date of Event
March 18, 2014
Report Date
March 18, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICAL SUPPORT FOUND THE NURSE CALL WAS MOST LIKELY NOT WORKING DUE TO THE UNIVERSAL TELEVISION POWER CONTROL BOARD HAVING A RELAY CLICK. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE HILL-ROM TECHNICAL SUPPORT ADVISED THE CUSTOMER TO SWITCH UNIVERSAL TELEVISION POWER CONTROL BOARD TO RESOLVE THE ISSUE. THREE ATTEMPTS HAVE BEEN MADE REGARDING A RESOLUTION TO THIS CONTACT LINE, WITH NO RESPONSE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM TECHNICAL SUPPORT RECEIVED A REPORT FROM THE ACCOUNT STATING THE NURSE CALL WAS NOT WORKING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
230998 ADVANTA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment