HOMECHOICE
Report
- Report Number
- 1416980-2014-22521
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- May 7, 2014
- Report Date
- June 19, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. THE CAUSE OF THE IIPV EVENT WAS DETERMINED TO BE A USE ERROR FURTHER SPECIFIED AS THE TOTAL TIDAL ULTRA FILTRATION (UF) REMOVAL WAS SET TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2014 AT 20:30:39. DURING NIGHT DRAIN CYCLE 12, THE PATIENT'S ULTRAFILTRATION READING WAS 1745ML, INDICATING THE HOME PATIENT (HP) DRAINED 995ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 410503 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |