FDA Adverse Event Malfunction Summary report: N

ECHELON LINEAR CUTTER

MDR report key: 3931661 · Received July 14, 2014

Report

Report Number
3005075853-2014-04904
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 25, 2014
Report Date
June 26, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K110385
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE FOLLOWING INFORMATION WAS REQUESTED, BUT UNAVAILABLE: WHAT WAS THE APPEARANCE OF THE STAPLE AND CUT LINE RELATED TO THE ¿CARTRIDGE MISFIRE?¿ DID THE POST-OPERATIVE CARE OF THE PATIENT CHANGE BASED ON THE EVENT? WERE THERE ANY PATIENT CONSEQUENCES? ARE THERE PHOTOS AVAILABLE? IF YES, PLEASE SEND TO [email protected].

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: ONE PIECE SLED, DRIVERS, CARTRIDGE. THE ANALYSIS FOUND THAT ONE PLE60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH AN ECR60G LOADED IN THE DEVICE. ADDITIONALLY, THE RELOAD WAS RECEIVED PARTIALLY FIRED 1/3. UPON EVALUATION OF THE RELOAD, THE CARTRIDGE BODY, ONE PIECE SLED AND SOME DRIVERS WERE NOTED TO BE DAMAGED. NO OBVIOUS DAMAGE TO THE CARTRIDGE DECK WAS NOTED, WHICH SUGGESTS THE CARTRIDGE, MAY NOT HAVE BEEN FIRED OVER A HARD OBJECT. AS ADDITIONAL TESTING, THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SLEEVE GASTRECTOMY PROCEDURE, THE "CARTRIDGE MISFIRED." THE CUSTOMER SAID THE SLED LOOKS LIKE IT WAS "FRACTURED." THE CASE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. PATIENT CONSEQUENCE UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411122 ECHELON LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 RELOAD - ECR60G