FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 3931640
·
Received July 14, 2014
Report
- Report Number
- 3004753838-2014-04960
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- June 2, 2014
- Report Date
- June 16, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
(B)(4). ADDITIONAL INFORMATION/DEVICE EVALUATION. EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED NO AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER.
Description of Event or Problem · 1
PATIENT'S HUSBAND CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014, TO CLAIM NO AUDIO OUTPUT ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT'S HUSBAND TO TEST THE ALERT FUNCTIONALITY AND REPORTED ALERTS WERE NOT FUNCTIONAL. DEXCOM HAS ISSUED AN RGA FOR PATIENT'S HUSBAND TO RETURN THE DEVICE TO BE INVESTIGATED
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 410247 | G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Other |