FDA Adverse Event Injury Summary report: N

M2A-MAGNUM PF CUP 48ODX42ID

MDR report key: 3931586 · Received July 14, 2014

Report

Report Number
0001825034-2014-06144
Event Type
Injury
Date Received
July 14, 2014
Date of Event
June 13, 2014
Report Date
September 15, 2014
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: "DISLOCATION AND SUBLUXATION DUE TO INADEQUATE FIXATION AND IMPROPER POSITIONING. MUSCLE AND FIBROUS TISSUE LAXITY MAY ALSO CONTRIBUTE TO THESE CONDITIONS." THIS REPORT IS NUMBER 2 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-06143 & 06144).

Additional Manufacturer Narrative · 1

THE RETURNED HEAD AND CUP WERE EVALUATED AND WAS FOUND TO HAVE BEEN MANUFACTURED WITHIN SPECIFICATIONS. VISUAL EXAMINATION OF HEAD AND CUP APPEARED TO SHOW EVIDENCE OF DAMAGE CONSISTENT WITH THE REPORTED CUP DISLOCATION. THE HEAD AND CUP APPEARED TO SHOW EVIDENCE OF SUBLUXATION OF THE JOINT, SCRATCHING OF THE BEARING SURFACES AND POSSIBLE METALLIC TRANSFER MARKS. ALTHOUGH IT APPEARS THAT SOME TYPE OF THIRD BODY PARTICLE WAS PRESENT, THE SOURCE AND IDENTITY OF THIS AGENT COULD NOT BE ESTABLISHED DURING VISUAL EXAMINATION. EVALUATION WAS UNABLE TO IDENTIFY THE ROOT CAUSE FOR DISLOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT A TOTAL HIP ARTHROPLASTY ON (B)(6) 2009. SUBSEQUENTLY, A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2014 DUE TO DISLOCATION. THE ACETABULAR CUP AND MODULAR HEAD WERE REMOVED AND THE STEM REMAINED IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409872 M2A-MAGNUM PF CUP 48ODX42ID PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 256520

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| R