FDA Adverse Event Malfunction Summary report: N

ADVANCE BED

MDR report key: 3931548 · Received April 14, 2014

Report

Report Number
1824206-2014-01181
Event Type
Malfunction
Date Received
April 14, 2014
Date of Event
March 17, 2014
Report Date
March 17, 2014
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BED EXIT TAPE SENSOR STRIP INOPERATIVE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2006-2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE BED EXIT TAPE SENSOR STRIP TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED EXIT LIGHT WOULD TURN ON BUT WOULD NOT ALARM. THE BED WAS LOCATED AT THE ACCOUNT IN THE BED SHOP. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
226625 ADVANCE BED AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1155

Patients

Seq Age Sex Outcome Treatment
1