FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3931541 · Received July 14, 2014

Report

Report Number
1416980-2014-22511
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 18, 2014
Report Date
June 19, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. THIS COMPLAINT IS FOR A REPORT OF A USE ERROR, WHERE THE PATIENT CONNECTED TO THE PATIENT LINE DURING THE PRIME PHASE. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. SECTION ¿WARNINGS AND CAUTIONS¿ WARNS THE USER NOT TO CONNECT TO YOUR PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. SECTION ¿PREPARE FOR THERAPY¿ WARNS THE USER THAT CONNECTING YOURSELF BEFORE "CONNECT YOURSELF" CAN CAUSE AIR TO BE DELIVERED TO THE PERITONEAL CAVITY. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT FAILED TO FOLLOW THERAPY STEPS DURING PRIME OF PERITONEAL DIALYSIS THERAPY. THE PATIENT WAS CONNECTED TO THE PATIENT LINE OF THE HOMECHOICE AUTOMATED PD SET WITH CASSETTE DURING PRIMING. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE PATIENT IN ENDING THERAPY AND ADVISED THE PATIENT TO RESTART WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409822 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 48 YR HOMECHOICE