GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00606
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- May 28, 2014
- Report Date
- May 28, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER MODEL #0010-2400; LOT #401A; SERIAL #(B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITS MILD DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITS VERY MILD DETRITUS ADHESION. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER. (B)(4) - REFERS TO FORWARD FIRING OF THE SIDE FIRING SURGICAL FIBER; A CODE REQUEST HAS BEEN SUBMITTED.
IT WAS REPORTED THAT SIGNIFICANT DIMINISHED VAPORIZATION EFFICIENCY WAS OBSERVED AT 81,078 JOULES. THE PROCEDURE WAS COMPLETED USING A SECOND FIBER AT 14,156 JOULES. PATIENT OUTCOME "OK" WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 409906 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 401A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |