FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3931029 · Received July 14, 2014

Report

Report Number
2955842-2014-04240
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
May 26, 2014
Report Date
June 17, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION OBSERVED THAT THE PITCH DOWN CABLE WERE FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT THE WRIST WERE UNDAMAGED. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S MAIN TUBE EXHIBITED SCRATCH MARKS/ABRASIONS. THE DISTAL END OF MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .069 - .289 AND NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S PITCH CABLE AND/OR MAIN TUBE COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI PROSTATECTOMY PROCEDURE, THE CUSTOMER NOTED THAT THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT CABLE WAS BROKEN. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410413 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10131017 196

Patients

Seq Age Sex Outcome Treatment
1