FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3930678 · Received July 12, 2014

Report

Report Number
2032227-2014-03716
Event Type
Malfunction
Date Received
July 12, 2014
Date of Event
June 11, 2014
Report Date
June 11, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED DURING TESTING. NO UNEXPECTED SCROLLING DURING TESTING. CRACKED CAST AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW, MISSING END CAP STICKER, CRACKED LCD WINDOW AND CRACKED BELT CLIP SLOT.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 78 MG/DL. THE BUTTONS WERE UNRESPONSIVE WHEN SHE WAS TRYING TO PROGRAM A BOLUS. NO SIGNIFICANT EVENTS THAT LED TO THE EVENT. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP. NO FURTHER INFORMATION TO REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
408772 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 33 YR