PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3930675 · Received July 12, 2014

Report

Report Number
2032227-2014-03742
Event Type
Malfunction
Date Received
July 12, 2014
Date of Event
June 11, 2014
Report Date
June 11, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. INSULIN PUMP UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. NO MOTOR ERROR ALARM NOTED, MOTOR PASSED MOTOR TEST. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED RESERVOIR TUBE LIP, CRACKED BELT CLIP SLOT, AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED. CUSTOMER STATED HE WAS ABLE TO REWIND THE PUMP. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND TO REVERT TO BACK-UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
408771 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 44 YR