FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3930614
·
Received July 12, 2014
Report
- Report Number
- 3004209178-2014-86826
- Event Type
- Malfunction
- Date Received
- July 12, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED DISPLACEMENT TEST, OCCLUSION, REWIND, BASIC OCCLUSION, PRIME AND NO DELIVERY TESTS. NO MOTOR ERROR ALARM OR NO DELIVERY ALARM NOTED. MOTOR PASSED MOTOR TEST. INSULIN PUMP HAD MINOR SCRATCHED DISPLAY WINDOW.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING A MOTOR ERROR AND NO DELIVERY ALARMS DURING BOLUS. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 256 MG/DL. CUSTOMER DECLINED TROUBLESHOOTING. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 408822 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |