FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3930591 · Received July 12, 2014

Report

Report Number
3004209178-2014-86759
Event Type
Malfunction
Date Received
July 12, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION. ALL BUTTONS FUNCTION PROPERLY. HOWEVER MOISTURE DAMAGE ON KEYPAD TRACES NOTED.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ON THE INSULIN PUMP WERE NOT RESPONDING. CUSTOMER STATED THAT THE INSULIN ALSO ALARMED AFTER BATTERY CHANGE. CUSTOMER WAS UNABLE TO CLEAR THE ALARM DUE TO THE BUTTONS NOT RESPONDING. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 16 MG/DL. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
408950 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWB

Patients

Seq Age Sex Outcome Treatment
1