FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3930517 · Received July 12, 2014

Report

Report Number
3004209178-2014-86705
Event Type
Malfunction
Date Received
July 12, 2014
Date of Event
May 11, 2014
Report Date
June 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

NO UNEXPECTED RESTART ERROR ALARM OR HISTORY ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP PASSED THE UNEXPECTED RESTART ERROR TEST AND SELF TEST. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD REPEATED ERROR ALARMS AND THE ACTIVATE BUTTONS WERE NOT RESPONDING. THE BLOOD GLUCOSE WAS UNKNOWN AT TIME OF CALL. NO FURTHER INFORMATION WAS PROVIDED.