FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3930453 · Received July 11, 2014

Report

Report Number
2531779-2014-19807
Event Type
Malfunction
Date Received
July 11, 2014
Report Date
July 4, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/22/2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT WAS VERIFIED IN THE HISTORY, BUT COULD NOT BE DUPLICATED DURING THE INVESTIGATION. UNEXPLAINED POR OBSERVED IN THE BLACK BOX ON (B)(6) 2014. NO PHYSICAL DAMAGE FOUND TO THE RETURNED BATTERY CAP OR THE BATTERY COMPARTMENT. RETURNED BATTERY CAP WAS ABLE TO FULLY TIGHTEN TO THE PUMP WITH NO YELLOW O RING SHOWING. PUMP WAS EXERCISED FOR 24HRS WITH RETURNED BATTERY CAP, NO POWER INTERRUPTIONS WERE DUPLICATED. THE RETURNED BATTERY CAP CONTACT MEASUREMENTS ARE WITHIN SPECIFICATIONS. THE PUMP COVER WAS REMOVED; NO EVIDENCE OF INTERNAL MOISTURE OR DAMAGE TO THE POWER CIRCUIT WAS FOUND.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
408114 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 12 YR