HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-22419
- Event Type
- Malfunction
- Date Received
- July 11, 2014
- Date of Event
- June 16, 2014
- Report Date
- June 16, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, IT WAS REPORTED THAT THE PATIENT WAS CONNECTED TO THE PATIENT LINE DURING THE PRIMING PHASE OF PERITONEAL DIALYSIS THERAPY. THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE, PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND SAYS TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
DURING TROUBLESHOOTING FOR A CHECK LINES AND BAGS ALARM, IT WAS REPORTED THAT THE PATIENT WAS CONNECTED TO THE PATIENT LINE DURING THE PRIMING PHASE OF PERITONEAL DIALYSIS THERAPY. THE CAREGIVER WAS INSTRUCTED TO DISCONNECT THE PATIENT AND CYCLE THE POWER OF THE HOMECHOICE DEVICE. PROPER PROCEDURES WERE REVIEWED AND THE CAREGIVER WAS ADVISED TO START THERAPY OVER USING ALL NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 406264 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | HOMECHOICE |