FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3930205 · Received July 11, 2014

Report

Report Number
1416980-2014-22414
Event Type
Malfunction
Date Received
July 11, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS A REPORT OF A USE ERROR, WHERE THE PATIENT CONNECTED TO THE SETUP BEFORE THE PATIENT LINE WAS FULLY PRIMED. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND WARNS THE USER NOT TO CONNECT TO YOUR PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. THE GUIDE ALSO INSTRUCTS THE USER TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED AND PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY REPRIMING THE PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING A REQUEST FOR THERAPY ASSISTANCE ON THE HOMECHOICE (HC) MACHINE, IT WAS REPORTED THAT THE HOME PATIENT (HP) HAD CONNECTED TO THE PATIENT LINE OF AN INTEGRATED APD SET WITH CASSETTE BEFORE IT WAS COMPLETELY PRIMED. THE CARE GIVER (CG) STATED THAT THEY THOUGHT PRIMING WAS COMPLETED, SO THE HP CONNECTED AND INITIATED THERAPY, BUT THE HC DEVICE STARTED PRIMING AGAIN. THE CG STATED THAT THEY NOTICED THIS IMMEDIATELY AND HAD THE HP RESTART THERAPY. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED PROPER PROCEDURES WITH THE CG. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
406460 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 52 YR HOMECHOICE