DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2014-00064
- Event Type
- Malfunction
- Date Received
- July 11, 2014
- Date of Event
- June 28, 2014
- Report Date
- June 29, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- MMI
- PMA / PMN Number
- K963498
- Removal / Correction Number
- 2517506-07/15/2014-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ORIGINAL MDR WAS SUBMITTED 07-11-2014. SIEMENS HAS CONFIRMED BACTERIAL CONTAMINATION IN CHEMISTRY WASH LOT RD41041. A POSITIVE BIAS WILL BE OBSERVED ON CTNI, LTNI, TSH, PBNP AND LPBN AND A NEGATIVE BIAS WILL BE SEEN ON FT4 WHEN THIS LOT IS FIRST PLACED ON THE INSTRUMENT WITHOUT RECALIBRATION OF METHODS. RECALIBRATION OF THESE ASSAYS WHEN USING THIS LOT WILL NOT CORRECT THE PROBLEM AND MAY CAUSE FALSELY HIGH OR LOW RESULTS. SIEMENS ISSUED AN URGENT FIELD SAFETY NOTICE DATED JULY 2014, COMMUNICATION #14-48, TO ALL ACCOUNTS WHO HAD RECEIVED THE IMPACTED LOT. CUSTOMERS WERE ADVISED OF THE IMMEDIATE REMEDIAL ACTIONS TO BE TAKEN BY THE CUSTOMER: IMMEDIATELY DISCONTINUE USE AND DISCARD ALL INVENTORY OF CHEMISTRY WASH LOT RD41041. IF THIS LOT WAS NEVER IN USE ON THE INSTRUMENT, NO FURTHER ACTION IS NEEDED. IF THIS LOT HAD BEEN IN USE: PLACE AN ALTERNATE LOT OF CHEMISTRY WASH ON THE DIMENSION SYSTEM. PRIME THE NEW BOTTLE OF CHEMISTRY WASH 30 TIMES. RECALIBRATE THE CTNI, LTNI, TSH, FT4, PBNP AND/OR LPBN ASSAYS USING THE ALTERNATE LOT OF CHEMISTRY WASH ONBOARD. FOR THE LISTED ASSAYS, PROCESS QUALITY CONTROL EVERY 8 HOURS. SIEMENS IS SHIPPING REPLACEMENT PRODUCT AT NO CHARGE FOR ALL ORDERS FILLED WITH LOT RD41041. AS AN ADDITIONAL PRECAUTION, A SIEMENS CUSTOMER SERVICE ENGINEER WILL BE MAKING A VISIT TO DECONTAMINATE THE SYSTEM(S) OF CUSTOMERS WHO HAVE USED LOT RD41041.
SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED COMPLAINTS OF DIMENSION(R) CHEMISTRY WASH LOT RD41041 CAUSING QC AND PATIENT SAMPLE RESULT SHIFTS. INTERNAL INVESTIGATION HAS CONFIRMED SHIFTS CAN BE ENCOUNTERED IN TWO SCENARIOS: 1.WHEN AN AFFECTED LOT OF CHEMISTRY WASH IS PLACED ON THE INSTRUMENT FOLLOWING AN UNAFFECTED LOT WITHOUT RECALIBRATION OF THE METHODS, A POSITIVE BIAS WILL BE SEEN ON CTNI, LTNI, TSH, PBNP AND LPBN. A NEGATIVE BIAS WILL BE SEEN ON FT4. 2.IF ANY OF THESE ASSAYS ARE CALIBRATED USING AN AFFECTED LOT OF CHEMISTRY WASH, QC SHOULD RETURN WITHIN RANGE; BUT WHEN A SUBSEQUENT UNAFFECTED CHEMISTRY WASH LOT IS PLACED INTO USE WITHOUT RECALIBRATION OF THE METHODS, A NEGATIVE BIAS WILL BE SEEN ON CTNI, LTNI, TSH, PBNP AND LPBN. A POSITIVE BIAS WILL BE SEEN ON FT4. SIEMENS HEALTHCARE DIAGNOSTICS INC. HAS AN ONGOING INVESTIGATION ON CHEM WASH.
FALSELY ELEVATED TROPONIN I AND PBNP RESULTS WERE OBTAINED ON QC SAMPLES ON THE DIMENSION INSTRUMENT. VALUES WERE OUTSIDE LABORATORY RANGES AND NO PATIENT RESULTS WERE REPORTED TO PHYSICIANS. AFTER A CHANGE TO AN ALTERNATE LOT OF CHEMISTRY WASH, QC RESULTS RETURNED TO WITHIN LABORATORY RANGES. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE FALSELY ELEVATED TROPONIN I AND PBNP QC RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I AND PBNP QC RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 406254 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CHEM WASH | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | RD41041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |