FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3930143 · Received July 11, 2014

Report

Report Number
2531779-2014-19789
Event Type
Malfunction
Date Received
July 11, 2014
Report Date
July 5, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 09/10/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: DURING A VISUAL INSPECTION OF THE PUMP, MOISTURE WAS OBSERVED BEHIND THE DISPLAY LENS. DURING TESTING, THE PUMP POWERED ON WITH APPROPRIATE AUDIBLE TONES; HOWEVER, THE DISPLAY SCREEN REMAINED BLANK AND NO VIBRATORY ALERTS WERE EMITTED. THE BLACK BOX DATA COULD NOT BE DOWNLOADED FOR REVIEW DUE TO MOISTURE DAMAGE. FURTHER ADEQUATE INVESTIGATION COULD NOT BE COMPLETED DUE TO ISSUES RELATED TO MOISTURE INGRESS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 052/053/054/055 SLEEP) ISSUE. IT WAS ALLEGED THAT THE ALARM HAD EMITTED THREE TIMES IN THE PAST THIRTY DAYS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
406087 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 39 YR