ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19785
- Event Type
- Malfunction
- Date Received
- July 11, 2014
- Report Date
- July 5, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/28/2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THERE WAS VISIBLE EVIDENCE OF MOISTURE BEHIND THE DISPLAY LENS. DUE TO THE MOISTURE CONTAMINATION, THE PUMP POWERED ON TO A DISTORTED DISPLAY SCREEN WITH FUNCTIONING AUDITORY AND VIBRATORY FEATURES; A POWER LOSS WAS NOT OBSERVED. THE BATTERY CAP WAS ABLE TO FULLY SECURE ONTO THE PUMP, AND THERE WAS EVIDENCE OF MOISTURE CONTAMINATION ON THE INTERIOR OF THE BATTERY CAP. THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. THE LEAK TEST FAILED DUE TO A BATTERY COMPARTMENT CRACK. THE KEYPAD BUTTONS WERE UNRESPONSIVE DUE TO INTERNAL MOISTURE DAMAGE. THE KEYPAD COVER WAS REMOVED AND NO CONTAMINATION WAS FOUND UNDER ANY OF THE KEYPAD BUTTON CONTACTS. THE PUMP WAS OPENED AND EVIDENCE OF MOISTURE WAS FOUND ON THE INTERIOR COMPONENTS OF THE PUMP. THE PUMP WAS OPENED AND EVIDENCE OF MOISTURE WAS FOUND ON ALL INTERNAL COMPONENTS OF THE PUMP.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. IT WAS ALLEGED THAT THERE WAS MOISTURE VISIBLE BEHIND THE DISPLAY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 408006 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |